A blank stores the optical contribution of your sterile media so the reading reflects biomass, not the medium.

What a blank is

A blank is an optical offset captured on sterile media before inoculation. The probe subtracts it from later readings so the signal reflects the cells you are growing rather than the background scattering of the medium itself.

When you need one

Capture a blank only when your media contains suspended solids that scatter light, such as complex or particulate media. For clear media, the background is negligible and no blank is needed.

Capturing a blank

  1. Prepare the sterile, thermally equilibrated medium before inoculation.
  2. Open the Sensor Unit's WiFi dashboard. If a measurement job is running, select Stop Job first.
  3. Under Job Control, select Take Blank.
  4. Wait for the capture to finish, then confirm that Blank in the References section shows a stored value.

Taking a blank replaces the previously stored blank. Capture a fresh one when the medium changes. If a blank from another medium is already stored before a clear-media run, take a new blank on the current medium so the old offset is not reused.

You can also capture a blank with a host command. See Run a measurement job for the complete web dashboard workflow.

Reference

The blank feeds the normalized_reflectance register; see the I²C interface reference. To convert readings into your own units, see Calibration.