Laser safety

The Turbid Vision Probe measures with an invisible near-infrared laser. It is designed to be eye-safe under normal use.

Designed to be eye-safe under normal use

The Turbid Vision Probe measures with an invisible near-infrared (NIR) laser, emitted as a diverging beam from an optical fiber into the culture. The laser is interlocked: it cannot turn on unless the Fiber Optic Probe is installed in its fitting, so the beam is terminated into the vessel during operation and is disabled whenever the Fiber Optic Probe is disconnected from the Sensor Unit.

At near-infrared wavelengths the cornea and lens absorb most of the energy before it reaches the retina, so the accessible-emission limits are much higher than for visible light. Combined with the diverging fiber output and the install interlock, the Turbid Vision Probe is designed to be eye-safe in normal operation, consistent with Class 1 under IEC 60825-1. As a routine precaution, do not stare into the optical aperture or view the output directly through optical instruments such as loupes, microscopes, or fiber scopes.

Safe handling

  • Keep the Fiber Optic Probe installed in its fitting during operation so the beam stays terminated into the vessel.
  • Do not stare into the optical aperture or view the output through magnifying optical instruments.
  • The laser disables automatically when the probe is disconnected from the Sensor Unit.

For the wiring and power context in which the laser operates, see Wiring connection.