Growth rate model
An onboard growth model estimates specific growth rate (µ) live and separates the biological signal from process noise.
What it does
The probe includes an onboard growth model, a Kalman Filter, that estimates the specific growth rate µ (in 1/h) continuously from the density signal. Because it models the underlying growth, it also produces a smoother, filtered density and rejects short-term process noise.
Enabling the model
The Kalman-filter-dependent registers return NaN until the model is enabled. Enable it from the web dashboard or over I²C by writing control command 0x08 to register 0x40 (see the I²C interface reference). The setting is persisted, so it survives a power cycle. Enable it while the measurement job is stopped: the command is rejected during a run, and it takes effect on the next job start. Once enabled, the following fields become available:
growth_rate(0x13): specific growth rate µ (1/h);filtered_calibrated_density(0x16): Kalman-filtered density in g/L, more stable than the unfiltered value;filtered_calibrated_growth_rate(0x17) andcalibrated_absolute_growth_rate(0x18): volumetric growth rates in g/L/h.
Filtered density and growth rate also require a loaded calibration, since they are derived from calibrated density. See Calibration.
Tuning
The Kalman filter tuning parameters are modifiable so you can match the filter to your process dynamics, trading responsiveness against smoothing. Adjust them from the dashboard. The full register list is in the I²C interface reference.