Skip to main content
Reacgen Biosystems

Pioreactor integration

The current early access Pioreactor integration path uses the probe's Stemma QT I²C interface. This page covers the Pioreactor-specific bits and links out for the shared wiring and register details while the installable plugin package is prepared.

Current status

Pioreactor can consume probe readings through the Stemma QT adapter. The integration is currently documented as an early access setup; an installable Pioreactor plugin package has not been published yet.

Connection

Wiring, power, and addressing are the same as any other host. See Wiring connection for the Qwiic pinout, the 3.3 V power rules, the onboard pull-ups, and how to change the I²C address. Pioreactor reads the probe over its Stemma QT I²C bus using repeated-START transactions.

What to publish

The fields most relevant to a Pioreactor job are calibrated_density (0x12) or the Kalman-filtered filtered_calibrated_density (0x16) in g/L, growth_rate (0x13) in 1/h, and reading_number (0x30) to detect fresh data. The full register map, status bits, and error flags are in the I²C interface reference.

Startup and staleness

Follow the start-job and first-reading sequence in the First run (start the job, poll job_running then has_initial, and allow at least 180 seconds for the 60-sample initial reference). On the host side, poll reading_number and only publish a new 0x12 or 0x16 value when the counter changes, so you never republish a stale reading. Treat NaN as unavailable: density needs a calibration lookup table, and the Kalman-filtered fields need the Kalman filter enabled.

Plugin status

The repository contains an integration design guide, but it does not contain a distributable Pioreactor plugin package. This page will link the package after it is published.