Performance
What the probe measures and how well.
Measurement principle
The probe measures near-infrared (NIR) diffuse reflectance. An invisible NIR laser is delivered through the fiber into the culture, and the backscattered light is collected and compared against the emitted power to deduce a reflectance ratio. That raw ratio is normalized against an initial reference (and an optional blank), then mapped to biomass density in your own units through a multi-point calibration. The onboard growth model, a Kalman Filter, estimates specific growth rate live and separates the biological signal from process noise.
The measurement uses a tight NIR band and rejects interference from ambient light, photobioreactor grow lights, and color changes in the medium.
Reflectance vs absorbance
Benchtop optical density is an absorbance measurement: light passes through a sample and the instrument reads how much of it is lost on the way through. Because the transmitted light falls off steeply as cells accumulate, the usable signal decays quickly in a growing culture. Past roughly OD 0.5 almost all the light is extinguished, the reading flattens out, and the sample has to be diluted to bring it back into the linear range. That is why dense cultures measured on a spectrophotometer need repeated serial dilutions.
The Turbid Vision Probe measures diffuse reflectance instead: it reads the light the cells scatter back toward the probe, not the light they remove. Backscatter keeps climbing with cell concentration well beyond the point where a transmitted beam would be fully extinguished, so a single probe stays linear from low-turbidity inoculation to high-density culture, in situ, without dilution or sampling.
Performance specs
| Measured parameter | Total cell density / reflectance / turbidity |
| Measurement principle | Near-infrared (NIR) diffuse reflectance |
| Wavelength | Near-infrared (NIR), invisible |
| Measuring range | 0.01–300 g/L dry cell weight (S. cerevisiae); application-dependent |
| Growth-rate output | Specific growth rate µ (1/h) via onboard Kalman filter |
| Reported units | Dry cell weight (g/L), OD600, FNU, NTU, or custom (calibration-dependent) |
| Sterilization | Autoclavable Fiber Optic Probe (detachable) to 130 °C; cycle count under review |
| Process connection | Immersion, vessel-adapted |
Scope and limitations
The probe measures total biomass (total cell density), not viable-cell density. It is not a GMP-validated PAT instrument; qualification for regulated environments is your responsibility. Within its rated range it is built for continuous monitoring and process control.
The measurement assumes the density is constant throughout the vessel. For fast-settling cultures and suspended solids, this assumes the culture is continuously stirred or agitated. Continuous agitation, together with the high polish of the Fiber Optic Probe optical face, prevents sensor fouling.
Backscatter depends on cell size as well as cell concentration, so a calibration is specific to the organism it was built with. Cultures whose cells differ markedly in size should each be calibrated on their own, and a calibration should not be carried from one organism to another without re-checking it.
The measurement also assumes cells are dispersed evenly through the sampled volume. Heavy clumping for typically disperse cultures is usually also a sign of insufficient agitation; keeping the culture well mixed improves both the culture and the reading.
For the digital interface, see Electronics; for physical dimensions and mounting, see Mechanical.