Cleaning optical surfaces
Clean the fiber connector faces every time you connect the probe to the Sensor Unit.
Why clean at every connection
The Fiber Optic Probe connects to the Sensor Unit through optical faces that must be clean for an accurate reading. A protective cap keeps new contamination off, but it does not leave the face clean, so do not assume a capped face is ready to use. Clean the mating faces every time you connect the probe.
The optical faces
The system has five protective caps: two on the Sensor Unit optical ports, two on the Fiber Optic Probe connectors, and one on the Fiber Optic Probe tip. When you connect the probe, clean the four connector faces: the two probe connectors and the two Sensor Unit ports. The wetted tip is cleaned separately; see Care & handling.
What you'll need
- An FC (2.5 mm) click cleaning pen, or single-use fiber cleaning sticks.
Procedure
- Uncap a face only when you are ready to clean and connect it.
- Clean each of the four connector faces with the FC (2.5 mm) cleaning pen or a fresh single-use cleaning stick: the two Fiber Optic Probe connectors and the two Sensor Unit ports.
- Connect the probe to the Sensor Unit.
- Re-cap any face that will not be connected, and keep caps on during storage.
Use IPA only
Use isopropyl alcohol (IPA) only. Do not use acetone or other chemical solvents on any part of the probe or Sensor Unit.