Troubleshooting
Physical and measurement issues during operation, and how to clear them. For I²C, register, or bus problems, see Troubleshooting in the integration guide.
There is no signal, or the reading looks like clear media.
Confirm the Fiber Optic Probe tip is fully immersed and deep enough to stay submerged at your minimum working volume; a tip in the headspace or out of the liquid reads like clear media. Check the wetted tip length against your reactor, and keep the tip clear of the impeller and at least 30 mm from the reactor bottom. See Sizing the probe and Mounting the probe.
The laser does not turn on.
The laser is interlocked: it cannot turn on unless the Fiber Optic Probe is installed in its fitting, and it is disabled whenever the probe is disconnected from the Sensor Unit. Confirm the probe is seated in its fitting and connected to the Sensor Unit. See Laser safety.
The reading drifts over time with no biomass change.
Check whether the probe has reached thermal equilibrium with the culture, and inspect the optical face for fouling. Continuous agitation and the polished optical face limit fouling; cleaning and equilibration guidance still needs hardware validation, so follow the procedure supplied with your probe revision. See Maintenance.
Density readings are inaccurate.
Build or replace the calibration with standards from the actual organism and media, and include standards across your intended operating range, since the model extrapolates beyond the outer calibration points. For settling cultures and suspended solids, calibrate and measure with the vessel continuously stirred. See Calibration.
Nothing in this list matches my problem.
For I²C, register, or bus problems, see Troubleshooting in the integration guide. Otherwise email info@reacgen.com with your host hardware, firmware/software version, and a brief description.