Wiring connection

The sensor is a 3.3 V I²C peripheral on a Qwiic / Stemma QT connector.

Wiring

Use the supplied Qwiic / Stemma QT cable. It is keyed and carries both power and signal, so a single cable connects the probe to any compatible host.

PinColourSignal
1YellowSCL
2BlueSDA
3Red3.3 V
4BlackGND

The probe has onboard I²C pull-up resistors, so no external resistors are needed.

Power

The host must use 3.3 V logic. On a 5 V board (classic Uno/Nano) use a level shifter, or use a 3.3 V host such as a Raspberry Pi, ESP32, ESP8266, or Arduino Nano.

Laser safety

Designed to be eye-safe under normal use.

The Turbid Vision Probe measures with an invisible near-infrared (NIR) laser, emitted as a diverging beam from an optical fiber into the culture. The laser is interlocked: it cannot turn on unless the Fiber Optic Probe is installed in its fitting, so the beam is terminated into the vessel during operation and is disabled whenever the Fiber Optic Probe is disconnected from the Sensor Unit.

At near-infrared wavelengths the cornea and lens absorb most of the energy before it reaches the retina, so the accessible-emission limits are much higher than for visible light. Combined with the diverging fiber output and the install interlock, the Turbid Vision Probe is designed to be eye-safe in normal operation, consistent with Class 1 under IEC 60825-1. As a routine precaution, do not stare into the optical aperture or view the output directly through optical instruments such as loupes, microscopes, or fiber scopes.

I²C address

Default 7-bit address: 0x69. If it collides with another device, change it under Settings > Device in the probe web interface. The configurable range is 0x08 to 0x77, and saving a new address reboots the probe so the I²C peripheral restarts on it. Update your host configuration to match.

Host integration

Any host with a 3.3 V I²C bus can read the probe: Raspberry Pi, ESP32, PLC gateway hardware, or lab automation controllers. Reads use a repeated-START transaction (write the register address, no STOP, then read). The full transaction structure, register map, and status bits are in the I²C interface reference; ready-made client libraries can be found on the Code libraries page.

Choosing a measurement register

RegisterFieldValueUse
0x10measured_reflectanceUnitless raw reflectanceLow-level optical diagnostics.
0x11normalized_reflectanceUnitless normalized reflectanceUses the initial reference and the optional blank reference.
0x12calibrated_densityUnfiltered density in g/LUse this when a calibration lookup table is loaded (status bit 7 set).
0x16filtered_calibrated_densityKalman-filtered density in g/LSmoother monitoring when the Kalman filter is enabled.

Verify the connection

Scan the bus and confirm the probe ACKs at its configured address, then run the browser-based procedure in Run a measurement job. When you first connect the probe to the Sensor Unit, clean the optical connector faces; see Cleaning optical surfaces. If the device does not respond, reseat the cable and check that the host bus is 3.3 V before swapping hardware. More recovery steps are on the Troubleshooting page in the integration guide.