First run
Set up the Sensor Unit once, then mount, sterilize, connect, and take your first reading, in the order you will actually do them.
What you'll need
- A reactor with a port that seals around the 2 mm Fiber Optic Probe. See Mounting the probe.
- The Sensor Unit, the Fiber Optic Probe, and the Qwiic / Stemma QT cable.
- A 3.3 V I²C host to read the probe, or your control stack for logging and control.
- An FC (2.5 mm) click cleaning pen or single-use fiber cleaning sticks, and isopropyl alcohol (IPA) with lint-free wipes.
- Known-density standards if you plan to run a full calibration.
Step 1: First-time setup (once per Sensor Unit)
Connect the Sensor Unit to your host over the Qwiic / Stemma QT cable, which also powers it, and confirm it responds at its I²C address (default 0x69). For wiring, power, and addressing, see Wiring connection.
Connect the Sensor Unit to WiFi so it is ready to use and can be reached from the dashboard. See Connect to WiFi.
If you will log measurements or drive a control loop, set up your host integration ahead of time so it is ready when the run starts. See the Integration guide.
Step 2: Mount the Fiber Optic Probe
Remove the cap from the Fiber Optic Probe tip, then mount the probe through the reactor headplate so the tip is fully immersed at your minimum working volume and sits at least 30 mm from the reactor bottom and clear of the impeller, wall, and other in-vessel hardware. For fittings and insertion depth, see Mounting the probe and Sizing the probe.
Step 3: Autoclave the vessel
Autoclave the vessel with the Fiber Optic Probe mounted but detached from the Sensor Unit. Remove all three caps from the probe first: the tip cap and the two connector caps. Caps stay off during the cycle.
The Sensor Unit is not autoclavable. Keep it out of the autoclave, and leave its two optical port caps in place until you are ready to connect.
Never autoclave the Sensor Unit
Only the Fiber Optic Probe is autoclavable. The Sensor Unit houses the optics and electronics and must stay out of the autoclave.
Step 4: Clean the connectors and attach the Sensor Unit
After the vessel has cooled, clean the two Fiber Optic Probe connector faces and the two Sensor Unit port faces, then connect the probe to the Sensor Unit. Clean the optical faces every time you connect, even if the caps were on: a cap keeps new contamination off but does not leave the face clean. See Cleaning optical surfaces.
Step 5: Capture a blank (suspended-solids media only)
If your media contains suspended solids that scatter light, capture a blank on the sterile, equilibrated media before inoculation. The blank stores the optical offset from the media so the reading reflects biomass. If your media is clear, you do not need a blank. If the dashboard shows a blank from a different medium, capture a fresh blank on the current medium before you start recording. See Blank reference.
Step 6: Start measuring
Configure the run, optionally create an on-device log, start the measurement job, and confirm live data from the web dashboard. See Run a measurement job. To report biomass in your own units, load a calibration; see Calibration.