Care & handling
Cleaning, storing, and handling the probe between runs.
Handling the Fiber Optic Probe
The Fiber Optic Probe carries light through glass fiber, so treat it with care. Do not bend it sharply or coil it tighter than its minimum bend radius, which can crack or break the fiber and end the probe's life. Avoid pinching, crushing, or dropping the probe, and do not pull on the fiber to move or disconnect it.
The probe is robust and built to take the normal wear and tear of routine use, but it is still a precision optical part and should be handled delicately.
Cleaning the probe
Clean the Fiber Optic Probe after each run and before storage, while the tip is still wet. Rinse the wetted section, then wipe the tip with a lint-free wipe dampened with isopropyl alcohol (IPA). Use IPA only; no other chemical solvents are recommended on the probe.
To clean the fiber connector faces where the probe meets the Sensor Unit, see Cleaning optical surfaces.
Storage
Store the Fiber Optic Probe clean and dry with all caps in place: the tip cap, the two connector caps, and the two Sensor Unit port caps. Caps protect the optical faces during storage, but they do not clean them, so still clean the connector faces before the next connection. See Cleaning optical surfaces.