Run a measurement job
Set up, start, monitor, and stop a measurement job from the Sensor Unit's web dashboard.
Before you begin
Confirm that:
- the Fiber Optic Probe is mounted, immersed, and connected to the Sensor Unit;
- the Sensor Unit is powered and connected to WiFi;
- you can open
turbidvision.local, or the unit's custom<name>.localaddress, from a browser on the same network; - the dashboard shows no blocking condition.
Setting up the hardware for the first time?
Complete First run before using this guide. It covers mounting, sterilization, cleaning the optical connectors, and connecting the Sensor Unit.
Step 1: Configure the job
Open Settings. The shipped defaults are appropriate for a normal first run, so change only the settings required by your process. Settings are stored on the Sensor Unit and remain in effect after a restart.
| Setting | What it does | Starting point |
|---|---|---|
| Reading interval (ms) | Sets the pause between measurement cycles. Acquisition time is added to this interval, so it is not the exact time between completed readings. | Leave the default unless you need a slower measurement cadence. Use the separate log interval to reduce CSV size. |
| Rolling average window | Controls how many recent reflectance readings are included in the rolling average shown on the dashboard. | Leave the default. Increase it for a steadier reflectance trend; decrease it to show short changes more quickly. |
| Ambient update interval | Takes a new dark-background measurement every N readings so ambient light can be removed from the signal. | Leave the default for normal operation. |
| Capture an initial normalization reference | Averages the first readings of a run into a baseline for normalized reflectance. The reference is stored and reused until you clear it. | Turn it on when you want change from the start of the culture, or when running the Growth Model without a calibration. |
| Initial normalization samples | Sets how many readings are averaged into the initial reference. More samples produce a tighter baseline but delay normal measurement. | Start with the default. The dashboard shows progress while the reference is captured. |
| Auto-start the OD job | Starts measurement automatically after the Sensor Unit powers on. | Keep it off while establishing your workflow. Turn it on for an unattended installation only after references and logging are set up. |
Select Save changes in the bar at the bottom of the page. Leave Output power target, Thermal Compensation, and Growth Model tuning at their defaults for a normal run. Changing thermal compensation clears the stored blank and initial references, so both must be captured again.
Step 2: Set references and the Growth Model
Return to Dashboard and check the References section. A blank and an initial reference serve different purposes:
- Blank: use Take Blank on sterile, temperature-equilibrated media before inoculation when the media contains suspended solids. A blank is optional for clear media. The job must be stopped while the blank is captured. If a stored blank came from a different medium, taking a new blank replaces it. See Blank reference.
- Initial normalization: this is the baseline at the start of a culture. To capture a new one, stop the job, select Clear Initial, make sure initial capture is enabled under Settings > Startup, then start the job. If you do not clear it, the stored reference is reused.
The Growth Model smooths the density signal and estimates specific growth rate µ. Enable or disable it from Job Control while the job is stopped. With a loaded calibration it reports filtered density and growth rate. Without a calibration it reports normalized change from the initial reference and a reflectance trend, not density or a true density-based growth rate. In that uncalibrated mode, the Sensor Unit automatically captures an initial reference if none is stored. See Growth rate model.
Step 3: Set up on-device logging (optional)
- Open Logging.
- Enter a short, unique Log name, such as the reactor or batch ID.
- Set the log interval in seconds. Leave it empty to record every completed sensor reading, or use a longer interval to reduce file size.
- Select Create Log. If the job is stopped, the log shows Armed: waiting for job and begins recording when the job starts.
Reading interval and log interval are different
The reading interval controls how often the sensor measures. The log interval controls how often a completed reading is written to the device log. Changing the log interval does not change the measurement itself.
Step 4: Start the job
- Return to Dashboard and select Start Job.
- If a new initial reference is required, the state changes to Initializing. Wait for the progress counter to finish before treating the displayed value as normal run data.
- Confirm that the state changes to Running, the data freshness indicator updates, and the expected measurement appears.
- Check Conditions. Open any condition to see what happened and the recommended recovery action.
Step 5: Read and monitor the result
The main dashboard value depends on how the Sensor Unit is configured:
| Configuration | Dashboard output | Meaning |
|---|---|---|
| No calibration, Growth Model off | Reflectance | The direct optical signal, reported in ppm. |
| Calibration loaded, Growth Model off | Density | The current unfiltered value from the active calibration. |
| Calibration loaded, Growth Model on | Filtered density and growth rate µ | Smoothed density plus the estimated specific growth rate in 1/h. |
| No calibration, Growth Model on | Change from baseline and reflectance trend | Relative change from the initial reference. This is not a calibrated density measurement. |
Use the dashboard chart for a quick live view. Use the on-device log or your host integration when the run record must be retained outside the current browser session.
Step 6: Stop and save the run
- On Dashboard, select Stop Job. Confirm that the state changes to Idle. This stops acquisition and turns off the laser.
- Open Logging and select Finish Log when the batch is complete.
- Select View to inspect the stored trace or CSV to download the full-resolution data.
Stopping a job does not finish its log
A stopped job leaves the current log armed. Starting the job again appends to the same log until you select Finish Log. This is useful for a brief pause, but finish the log before beginning a different batch.
Running from a host instead
The web dashboard and the digital interfaces control the same measurement job. For automation, start and stop the job over I²C or WebSocket, watch the job and initialization status, and consume a value only when the reading number changes. The complete commands, status fields, and timing behavior are documented in the I²C interface reference and WebSocket API reference.