Mechanical
Sensor Unit and Fiber Optic Probe dimensions, and how to mount and size the probe for your reactor.
Sensor Unit
The Sensor Unit houses the optics and electronics and connects to the Fiber Optic Probe. Use the dimensions below to plan bench placement, mounting, and cable routing. All dimensions are in millimeters.

Fiber Optic Probe
| Construction | Detachable Fiber Optic Probe + external Sensor Unit |
| Fiber Optic Probe diameter (wetted) | 2 mm rod |
| Wetted materials | 316L stainless steel, silica glass, and bio-compatible epoxy |
| Optical face polish | N1 (0.3 µm) |
| Wetted tip length options | 50, 75, 100, 230 mm, or custom |
| Fiber Optic Probe length options | 50, 75, 100, 150, 200 cm, or custom |
Mounting the probe
The wetted section of the Fiber Optic Probe is a 2 mm diameter rod. It passes through the reactor headplate and is held by a compression fitting that clamps and seals around the Fiber Optic Probe, which also sets the insertion depth. Any port that seals cleanly around a 2 mm rod will work. The cross-section below shows a typical installation.

Mounting options
Off-the-shelf compression fitting
A 2 mm tube compression fitting seals around the Fiber Optic Probe and threads into a tapped port in your headplate. It comes in the common NPT and metric thread sizes below; pick the one that matches your port. This is the simplest option when your headplate has, or can be tapped with, a threaded port.
Before you mount
- Confirm the fitting and ferrule materials suit your process, including temperature and autoclaving.
- Set the insertion depth from your wetted tip length so the tip stays fully immersed at the minimum working volume.
- Keep the Fiber Optic Probe tip clear of obstructions and at least 30 mm from the reactor bottom. The probe reads a small volume of culture directly in front of the tip, so position it so that the vessel wall, impeller, baffles, sparger, and any other probes or tubes are not close to the tip, since those surfaces can reflect light back and bias the reading.
Working with a non-standard headplate or need a custom mount? Contact us at info@reacgen.com and we will help you find or design a mounting solution.
Mounting the Sensor Unit
By default the Sensor Unit simply rests on a bench or shelf. If you would rather fix it in place, for example to a panel, a rack, or the frame of your bioreactor, the Sensor Unit is also available in an optional mounting version. This version adds four mounting tabs to the enclosure, two on each side, each with a hole for an M3 screw.
The holes are spaced 80 mm apart across the width of the enclosure and 30 mm apart along its sides, measured center to center. Use the drawing below to lay out the holes on your mounting surface.

Sizing the probe
How to calculate the wetted tip length and the Fiber Optic Probe length for your reactor.
Wetted tip length
The wetted tip length sets how far the Fiber Optic Probe extends into the vessel. Select a length that keeps the tip fully immersed at your reactor's minimum working volume.
The wetted tip length is the sum of the three dimensions shown below (units in mm):
- M is the height from the top of your mount to the bottom of your reactor top plate.
- H is the maximum reactor headspace when filled to its minimum operating volume.
- D is the Fiber Optic Probe insertion depth into the liquid; we suggest a minimum of 20 mm.

Ensure the Fiber Optic Probe tip sits at least 30 mm from the reactor bottom, and away from the vessel wall, impeller, and other in-vessel hardware, so nothing sits in the small volume the probe reads directly in front of the tip. See Mounting the probe above for mounting options.
Fiber Optic Probe length
The Fiber Optic Probe length is dictated by where the Sensor Unit sits relative to the Fiber Optic Probe mounting position and the minimum fiber bend radius. Calculate it from the dimensions shown below (units in cm):
- W is the horizontal distance from the Sensor Unit to the mounting point.
- L is the vertical distance from the surface the Sensor Unit rests on to the reactor top plate.

This is the minimum probe length. Choose the next standard length up; any extra length simply makes the loop taller.