Skip to main content
Reacgen Biosystems

Troubleshooting

I²C, register, and bus issues, and how to clear them. For physical and measurement issues, see Troubleshooting in the operating manual.

The probe is not found on the I²C bus.

Reseat the cable, scan the bus, and confirm that the device responds at the default 7-bit address 0x69 or the alternate address configured under Settings > Device in the probe web interface. Verify 3.3 V power before swapping hardware.

reading_number is not changing.

The measurement job may be stopped or still initializing. Send CMD_START_JOB (0x01), poll status register 0x20 until job_running is set, then allow at least 180 seconds for has_initial. Compare register 0x30 between reads to detect fresh data.

A register returns NaN.

NaN means the requested value is unavailable. Start the job and wait for has_initial. calibrated_density also requires a loaded calibration lookup table. Kalman-filter-dependent registers require the Kalman filter to be enabled through the probe web interface or over I²C (control command 0x08).

The status byte shows an error (bit 6 set).

Status bit 6 is the OR of every system fault and warning. Read register 0x21 (uint32, little-endian) for the detailed vector and decode it against the error-flags table in the I²C reference. Error-class bits (0 to 4, e.g. PI_FAILED, STORAGE) latch and clear with CMD_CLEAR_ERRORS (0x07); warning-class bits (5 to 8, e.g. WIFI_DOWN, OTA) are live-derived and re-assert while their condition persists.

I have two probes conflicting on the I²C bus.

Open the probe web interface, change one probe's I²C slave address under Settings > Device, and update the host configuration. The valid configurable range is 0x08 to 0x77, and changing the address reboots the probe.

Nothing in this list matches my problem.

Email info@reacgen.com with the host hardware, firmware/software version, and a brief description.